Product information
Relative humidity control
HUM-1
Humidity generator provides constant humidity though feedback control by simply setting the target relative humidity (RH). It is suitable for generating humid gas at a relatively small flow rate of several hundred mL per minute, and is ideal for thermal stability testing, continuous measurements of hygroscopic and dehydration changes in a humid atmosphere.
HUM Features
- Type: Divided-flow type (single tank bubbling method), feedback control of dry gas flow by temperature and humidity sensors placed near the test sample
- Humidity stability: Within ±2% RH (excluding transient states during setting changes)
HUM Specifications
| Model name | HUM-1 |
|---|---|
| System | Humidity generator |
| Water vapor generation range | Room temperature to 80℃ dry to 90% RH |
| Wet gas flow rate | 200 to 500 mL/min |
| Transfer temperature | Max 110℃ |
| Water tank temperature | Room temperature to 90℃ |
| Dry gas flow rate | 0~500 mL/min |
| Variable stabilization time | Within 10 minutes (when variable 10% RH) |
| Continuous operating time | 40 hours at 80℃, 90%RH 100 hours at 60℃, 90%RH |
| Variable humidity method | Feedback type humidity control system |
| Humidity control method | PID control method |
| Humidity stability | Within ±2% RH |
| Humidity sensor | Polymer type relative humidity sensor |
| Temperature measurement element | Pt resistor elements |
| Gas type | Dry nitrogen |
| Bubbler capacity | 300 mL |
| Measurement atmosphere | Air, inert gas, water vapor |
| Dimensions | 290(W) x 260(H) x 350(D) mm |
| Mass | 15 kg |
| Power requirements | 1Ø, AC100 V, 50/60 Hz, 5 A |